ADM Flow Meter
Installing and Replacing the Cartridge
Agilent CrossLab CS ADM Flow Meter Operation Manual
31
Installing and Replacing the
Cartridge
Installing the cartridge
The CrossLab CS Mainframe is a platform that can accept the
ADM Flow Meter Cartridge P/N G6692A and Electronic Leak
Detector Cartridge P/N G6694A.
NOTE
If you purchased the leak detector cartridge standalone,
update your existing mainframe to the latest firmware when
using your leak detector for the first time.
Swapping or replacing the cartridge
with ADM flow meter cartridge
1
Press and hold the Power/Mode button for three seconds
to power down the unit.
2
Loosen both captive thumbscrews completley and lift the
cartridge out of the CrossLab CS Mainframe body.
Captive screws