Installing and uninstalling labware racks
G5580A BenchCel Microplate Handler Quick Guide
Uninstalling labware racks
To uninstall a rack when rack-release button is green:
1
Verify that the rack-release button on the stacker head is lit green, and then press it.
A click sounds as the locks retract.
2
While the green light is flashing, lift the labware rack up and out of the stacker
head.
Note:
If the green light stops flashing, press the rack-release button again.
To uninstall a rack when rack-release button is blue:
1
In
BenchCel Diagnostics
, initialize the profile.
2
In the
Controls
tab, click
Unload All
to unload all the stackers. See figure, item
1
.
In the Controls tab image: (
2
) A blue rack-release button indicates that the stack is
loaded. (
3
) A rack image appears on the stacker if the sensors detect a rack. (
4
) A
green labware image indicates that the sensors detect a plate.
3
To unlock the rack, press the green rack-release button at the top of the stacker
head. Wait for the click to sound as the locks retract. While the green light is
flashing, lift the labware rack up and out of the stacker head.