d
a
COMMAND
ERROR
DETECTED
-109
"Missing
parameter"
Return
to
Execute
GPIB
command
230
A=SPOLL(@Hp4396)
240
OUTPUT
@Hp4396;"*ESR?"
250
ENTER
@Hp4396;Estat
260
ENABLE
INTR
7
!
When
iBASIC
is
used,
change
"7"
to
"8"
270
RETURN
Lines
230
to
270
clear
SRQ
before
returning
to
the
main
routine
.
Line
230
reads
the
analyzer's
status
byte
.
The
A=SPOLL(@Hp4396)
statement
reads
the
Status
Byte
Register
of
the
address
@Hp4396 (analyzer),
and
enters
the
value
into
A .
The
command
error
causes
the
SRQ
and
with
bit
5
and
bit
6
of
the
Status
Byte
Register
set,
the
value
of
A
is
96.
Reading
the
Status
Byte
Register
by
using
the
SPOLL
command
clears
SRQ
(status
byte
bit
6).
In
line
240
and
line
250,
the
command
*ESR?
reads
the
contents
of
the
Standard
Event
Status
Register
.
With
Bit
5
of
Standard
Event
Status
Register
set,
the
value
of
Estat
is
32.
Reading
the
Standard
Event
Status
Register
by
using
the
*ESR?
command
clears
the
register
.
A
branch
to
Err_report
disables
the
interrupt.
Therefore
,
the
return
from
Err_report
must
reenable
the
interrupt.
Synchronizing
the
Analyzer
from
Remote
3-7
Содержание Agilent 4396B
Страница 12: ...Tables A 1 Manual Changes by Serial Number A 1 A 2 Manual Changes by Firmware Version A 1 Contents 6 ...
Страница 32: ......
Страница 52: ......
Страница 68: ......
Страница 72: ......
Страница 92: ......
Страница 96: ......
Страница 99: ...Agilent 4396B Network Spectrum Impedance Analyzer Using Instrument BASIC with the 4396B Printed in JAPAN ...
Страница 100: ...c Copyright 1997 2000 2001 2003 Agilent Technologies Japan Ltd ...
Страница 116: ......
Страница 126: ......
Страница 130: ......
Страница 138: ......
Страница 148: ......
Страница 163: ...C The Keyboard Mini DIN keyboard is following key layout Figure C 1 mini DIN Keyboard The Keyboard C 1 ...
Страница 176: ......