Technical Information
Examples
17
Turbo-V 750/850-AG Rack Controller User Manual / 87-901-023-01
(
B
)
293/304
Auto Test Phase:
In stop status, it is possible to take the three phase test diagnosis
pressing the CO ME LOW SPEED buttons at the
same time for at least 2 sec.
This test checks the motor windings electrical continuity and
symmetry. Furthermore it is able to detect a short circuit between a
motor phase winding and the Protection Earth.
Figure 44
S
TART
/S
TOP
L
OW SPEED
C
OUNTERS
M