TECHNICAL INFORMATION
7
0 87-900-873-01 (
F
)
CONTROLLER SPECIFICATIONS
Input:
Voltage
Frequency
Power
100, 120, 220, 240 Vac
]
10%, 1-phase
47 to 63 Hz
600 VA maximum
Output:
Voltage
Frequency
Power
water cooling
air cooling
54 Vac nominal
]
10%
700 Hz
]
2%
420 W maximum
320 W
220 W
Operating temperature
0°C to +40 °C
Storage temperature
-20°C to +70°C
Fuse (mains)
2 x T4A(slow blow) for 220 or 240 input Voltage
2 x T8A(slow blow) for 100 or 120 input Voltage
P1 optoisolator input
Minimum ON 3mA Maximum 5mA
J2 optoisolator output
24 Vdc, 60 mA
J6 output (fan)
120Vac, 1A disregarding the mains
J7 output (forepump coil, etc.)
120 Vac, 1A
disregarding the mains
Radio interference suppression
EN 55011 class A group 1
EN 61010-1
IEC1000-4-2,1000-4-3, 1000-4-4
Auxiliary connectors
P1
J2
J6
J7
J14
External INPUT signals (pins)
OUTPUT signals (sockets)
Output fan Voltage
Valve, forepump socket
RS 232/422/485 connection (optional)
Interconnecting cables
Mains cable (3-wire, 3-meter long)
Pump cable (6-wire, 5-meter long)
Weight (both models)
10.7 kg (24 lbs)