TECHNICAL INFORMATION
8
9 87-900-873-01 (
F
)
•
Enter the value by pressing the PUMP CURRENT
push-button.
•
Press the CYCLE NUMBER push-button to select
the second number.
S P E E D
T H R E S H O L D
S E L E C T
I
O N
:
X X K R P M
•
Enter the value by pressing the PUMP CURRENT
push-button, and the display shows:
R U N
-
U P
T
I
M E
S E L
:
X X h
X X m
X X
s
where:
RUN-UP TIME
= is the interval time from start to
speed threshold value in hours, minutes, seconds.
Select from 00 to 99 hours, and from 00 to 59 minutes
or seconds.
Select the run-up time according to the chamber
volume and/or operating cycle feature (see the
preceding cycle diagram) by pressing the CYCLE
NUMBER push-button to select the desired number.