![Agilent Technologies 9130011 Скачать руководство пользователя страница 46](http://html1.mh-extra.com/html/agilent-technologies/9130011/9130011_user-manual_2868741046.webp)
4
Instructions for Use
General
Information
46/58
Miniature and 2 l/s VacIon Pumps User Manual 87-400-044-01 (F)
The complete pumping system consists of the VacIon pump, its
magnet, and a power supply. These VacIon pump is a pump which
operates on the principle of ion gettering in the pressure range from
10
-3
Torr to less than 10
-9
Torr. A roughing pump is needed to lower
the system pressure from atmosphere to below 10
-3
Torr.
The pump should be powered by a positive voltage of 3000 – 3500
Volt (open circuit), direct current supply and the short circuit
current of the control unit should match the current drawn by the
pump at its maximum operating pressure.