Chapter 1
1-47
Alphabetical Command Reference
CSWI
CSWI
Syntax
CSWI<ON|OFF>;
or
CSWI?;
NOTE
These commands only apply to ES models.
Description
See also
Front Panel Equivalents
Command
Description
Range
Query Response
CSWION
Selects test set continuous switching.
When continuous switching is on, the
analyzer measures all four S-parameters
each time before displaying the data for a
full 2-port cal measurement.
N/A
N/A
CSWIOFF
Selects test set hold mode. The analyzer
measures all four S-parameters once and
then measures the desired parameter
continuously. This is known as a fast
2-port cal measurement and it is less
accurate than a full 2-port calibrated
measurement.
N/A
N/A
CSWI?
Queries whether continuous switching is
on or off.
N/A
<0|1><
L
F
>
Command
Hardkey
Softkey
CSWION
CSWIOFF
Cal
TESTSET SW CONTINUOUS
Cal
TESTSET SW HOLD