![Agilent Technologies 8719ES Скачать руководство пользователя страница 312](http://html.mh-extra.com/html/agilent-technologies/8719es/8719es_programmers-manual_2867960312.webp)
Chapter 3
3-3
GPIB Programming
Analyzer Command Syntax
Some codes require appendages (
ON
,
OFF
,
1
,
2
, etc.). Codes that do not have a front-panel
equivalent are GPIB only commands. They use a similar convention based on the common
name of the function.
Valid Characters
The analyzer accepts the following ASCII characters:
• letters
• numbers
• decimal points
• +/
−
• semicolons (;)
• quotation marks (“)
• carriage returns (CR)
• linefeeds (LF)
Both upper- and lower-case letters are acceptable. Carriage returns, leading zeros, spaces,
and unnecessary terminators are ignored, except for those within a command or
appendage. If the analyzer does not recognize a character as appropriate, it generates a
syntax error message and recovers at the next terminator.
Table 3-1 Code Naming Convention
Convention
Key Title
For GPIB Code Use
Example
One word
First four letters
POWE
STAR
Two words
First three letters of first
word, first letter of second
word
ELED
SEAR
Two words in a
group
Four letters of both
MARKCENT
GATESPAN
Three words
First three Letters of first
word, first letter of second
word, first four letters of third
word
CALKN50
PENNDATA
Power
Start
ELECTRICAL DELAY
SEARCH RIGHT
MARKER
→
CENTER
GATE
→
SPAN
CAL KIT N 50
Ω
PEN NUM DATA
Содержание 8719ES
Страница 15: ...1 1 1 Alphabetical Command Reference ...
Страница 293: ...2 1 2 Introduction to Instrument Control ...
Страница 310: ...3 1 3 GPIB Programming ...
Страница 334: ...4 1 4 Reading Analyzer Data ...
Страница 343: ...5 1 5 Data Processing Chain ...
Страница 350: ...6 1 6 Error Reporting ...
Страница 364: ...7 1 7 Programming Examples ...
Страница 502: ...A 1 A Preset Conditions ...
Страница 517: ...B 1 B Command Listings ...