![Agilent Technologies 86060C Series Скачать руководство пользователя страница 72](http://html.mh-extra.com/html/agilent-technologies/86060c-series/86060c-series_user-manual_2868536072.webp)
4-22
Programming
Standard SCPI Commands
Standard SCPI Commands
:STATus:<node>:CONDition
The :STATus:<node>:CONDition query returns the value for the condition
register for the node. Condition registers have no function in this instrument,
but the query is included for compatability with the SCPI standard. This query
always returns the value 0.
Usage:
GPIB only
Query Syntax:
:STATus:<node>:CONDition?
Returned Format:
<value>
Where:
<value> = 0 (integer – NR1)
<node> = OPERation | QUEStionable
Example:
OUTPUT 711;":STATUS:OPERATION:CONDITION?"
ENTER 711;Value
PRINT Value
:STATus:<node>:ENABle
The :STATus:<node>:ENABle command sets the enable register for the node.
Enable registers have no function in this instrument, but the command is
included for compatability with the SCPI standard.
The :STATus:<node>:ENABle query returns the value of the enable register
for the node.
Usage:
GPIB only
Содержание 86060C Series
Страница 1: ...Agilent 86060C Series Lightwave Switches User s Guide ...
Страница 6: ......
Страница 17: ...1 11 General Information Specifications and Regulatory Information ...
Страница 32: ......
Страница 50: ......
Страница 79: ...4 29 Programming SYSTem CONFig Example DIM Config OUTPUT 711 SYSTem CONFIG ENTER 711 Config PRINT Config ...
Страница 91: ...5 Spare Channel Replacement Procedure 5 4 Electrostatic Discharge Information 5 7 Servicing ...
Страница 96: ...5 6 Servicing Spare Channel Replacement Procedure Figure 5 2 Top view of the Agilent 86062C lightwave switch ...
Страница 102: ......