4-15
Programming
*OPC (Operation Complete)
*OPC (Operation Complete)
The *OPC command sets the operation complete bit in the Standard Event
Status Register when all pending device operations have finished.
The *OPC query places an ASCII "1" in the output queue when all pending
device operations have finished.
Usage:
GPIB only
Command Syntax:
*OPC
Example:
OUTPUT 711;"*OPC"
Query Syntax:
*OPC?
Returned Format:
1<NL>
Example:
OUTPUT 711;"*OPC?"
ENTER 711;Op$
*RCL (Recall)
The *RCL command recalls the state of the instrument from the specified
instrument state register. If the instrument state register has not been previ-
ously stored, the *RCL command will restore the instrument to its power-on
state.
Usage:
GPIB and RS-232
NOTE
The *OPC command can be used to ensure all switch movement operations have com-
pleted before continuing the program. By following a ROUTE:LAYER:CHANNEL command
with an *OPC query and an ENTER statement, the program will pause until the response
(ASCII "1") is returned by the instrument.
Содержание 86060C Series
Страница 1: ...Agilent 86060C Series Lightwave Switches User s Guide ...
Страница 6: ......
Страница 17: ...1 11 General Information Specifications and Regulatory Information ...
Страница 32: ......
Страница 50: ......
Страница 79: ...4 29 Programming SYSTem CONFig Example DIM Config OUTPUT 711 SYSTem CONFIG ENTER 711 Config PRINT Config ...
Страница 91: ...5 Spare Channel Replacement Procedure 5 4 Electrostatic Discharge Information 5 7 Servicing ...
Страница 96: ...5 6 Servicing Spare Channel Replacement Procedure Figure 5 2 Top view of the Agilent 86062C lightwave switch ...
Страница 102: ......