![Agilent Technologies 85051B Скачать руководство пользователя страница 8](http://html.mh-extra.com/html/agilent-technologies/85051b/85051b_users-and-service-manual_2867929008.webp)
1- 2
85051B
General Information
Verification Kit Overview
Verification Kit Overview
The Agilent 85051B 7 mm verification kit provides a set of standards with known characteristics, traceable
to a reference (golden) standard in Agilent Technologies calibration lab. This set of standards is used to
verify your measurement calibration and also to verify that your PNA system is operating within its
specifications. The frequency range covered by the 85051B is from 45 MHz to 18 GHz.
Kit Contents
The 85051B verification kit includes the following items:
•
20 dB attenuator
•
50 dB attenuator
•
25
mismatch airline
•
50
airline
•
open-end wrench
•
storage box
•
User’s and Service Guide (this manual)
•
USB drive that contains factory-measured verification data for use with the PNA system verification
process
•
data sheet for each device that contains factory-measured verification data
•
anti-virus scan report
Refer to
Chapter 6, “Replaceable Parts,”
for a complete list of replaceable part numbers.
IMPORTANT
Backup copies of the data sheets, anti-virus scan reports, and the data contained on the USB
drive should be made immediately upon receipt of the verification kit.
NOTE
A file containing the verification data for your kit is maintained for one year from the time of
measurement. If you lose this data, contact Agilent. See
“Contacting Agilent” on page 5-5
Calibration Definitions
The 85051B verification kit is intended to be used with the 85050B/C/D 7 mm calibration kits. Prior to
performing a calibration with your PNA, the calibration kit must be selected and the calibration definitions
for the devices in the kit installed in the PNA. Refer to your PNA Help system for instructions on selecting
the calibration kit and performing a calibration.
Installation of the Calibration Definitions
The calibration definitions for the kit should be permanently installed in the internal memory or hard disk of
the PNA.
T
hey may already be resident within the analyzer, or you can download them from Agilent’s
Calibration Kit Definitions Web page at
www.na.tm.agilent.com/pna/caldefs/stddefs.html.
In addition
,
the
Содержание 85051B
Страница 4: ...iv 85051B ...
Страница 7: ...85051B 1 1 1 General Information ...
Страница 12: ...1 6 85051B General Information Clarifying the Terminology of a Connector Interface ...
Страница 13: ...85051B 2 1 2 Specifications ...
Страница 19: ...85051B 3 1 3 Use Maintenance and Care of the Devices ...
Страница 37: ...85051B 4 1 4 Performance Verification ...
Страница 40: ...4 4 85051B Performance Verification Recertification ...
Страница 41: ...85051B 5 1 5 Troubleshooting ...
Страница 43: ...85051B 5 3 Troubleshooting Troubleshooting Process Figure 5 1 Troubleshooting Flowchart ...
Страница 46: ...5 6 85051B Troubleshooting Contacting Agilent ...
Страница 47: ...85051B 6 1 6 Replaceable Parts ...
Страница 50: ...6 4 85051B Replaceable Parts Introduction Figure 6 1 Replaceable Parts ...