Good connections require a skilled operator.
The most common cause of measurement error is bad
connections.
The following procedures illustrate how to make good connections.
How to Make a Connection
Preliminary Connection
1. Ground yourself and all devices. Wear a grounded wrist strap and work on a grounded, conductive table
mat. Refer to
“Electrostatic Discharge” on page 3-2
for ESD precautions.
2. Visually inspect the connectors. Refer to
“Visual Inspection” on page 3-3
3. If necessary, clean the connectors. Refer to
“Cleaning Connectors” on page 3-4
4. Use a connector gage to verify that all center conductors are within the pin depth specifications in
“Gaging Connectors” on page 3-6
.
5. Fully extend the connector sleeve on one of the connectors. Spin its knurled connector nut to make sure
the threads are fully extended. Fully retract the sleeve on the other connector. The extended sleeve
creates a cylinder into which the other connector fits.
If one of the connectors is fixed (such as on a test port), fully extend that connector sleeve and fully
retract the sleeve on the moveable connector.
6. Carefully align the connectors. As you make the actual connection, be sure the connectors align
perfectly.
7. Push the connectors straight together. Do
not
twist or screw the connectors together.
8. Engage the connector nut (of the connector with the retracted sleeve) over the threads of the other
connector (the connector with the extended sleeve). Turn only the connector nut. Let the connector nut
pull the two connectors straight together.
9. Do not overtighten this connection. A connection in which the outer conductors make gentle contact at
all points on both mating surfaces is sufficient. Very light finger pressure is enough to accomplish this.
10. Make sure the connectors are properly supported. Relieve any side pressure on the connection from long
or heavy devices or cables.
Final Connection Using a Torque Wrench
Use a torque wrench to make a final connection.
provides information about the torque wrench
recommended for use with the calibration kit. A torque wrench is included in the calibration kit. Refer to
for part number and ordering information.
Table 3-1
Torque Wrench Information
Connector Type
Torque Setting
Torque Tolerance
7 mm
136 N-cm (12 in-lb)
13.6 N-cm (
1.2 in-lb)
Содержание 85051B
Страница 4: ...iv 85051B ...
Страница 7: ...85051B 1 1 1 General Information ...
Страница 12: ...1 6 85051B General Information Clarifying the Terminology of a Connector Interface ...
Страница 13: ...85051B 2 1 2 Specifications ...
Страница 19: ...85051B 3 1 3 Use Maintenance and Care of the Devices ...
Страница 37: ...85051B 4 1 4 Performance Verification ...
Страница 40: ...4 4 85051B Performance Verification Recertification ...
Страница 41: ...85051B 5 1 5 Troubleshooting ...
Страница 43: ...85051B 5 3 Troubleshooting Troubleshooting Process Figure 5 1 Troubleshooting Flowchart ...
Страница 46: ...5 6 85051B Troubleshooting Contacting Agilent ...
Страница 47: ...85051B 6 1 6 Replaceable Parts ...
Страница 50: ...6 4 85051B Replaceable Parts Introduction Figure 6 1 Replaceable Parts ...