![Agilent Technologies 85051B Скачать руководство пользователя страница 35](http://html.mh-extra.com/html/agilent-technologies/85051b/85051b_users-and-service-manual_2867929035.webp)
85051B
3-17
Use, Maintenance, and Care of the Devices
Making Connections
Figure 3-11
Torquing the Connection
How to Separate a Connection
NOTE
Do
not
turn the device body. Only turn the connector nut. Damage to the center conductor
can occur if the device body is rotated.
1. Use an open-end wrench to prevent the device body from turning.
2. Use another open-end wrench to loosen the connector nut.
3. Complete the disconnection by hand, turning only the connector nut.
4. Pull the connectors straight apart without twisting, rocking, or bending either of the connectors.
NOTE
If disconnecting an airline and the airline center conductor does not disengage from the
device center conductor, gently pull the center conductors apart and then push the airline
center conductor back inside the outer conductor of the airline.
Содержание 85051B
Страница 4: ...iv 85051B ...
Страница 7: ...85051B 1 1 1 General Information ...
Страница 12: ...1 6 85051B General Information Clarifying the Terminology of a Connector Interface ...
Страница 13: ...85051B 2 1 2 Specifications ...
Страница 19: ...85051B 3 1 3 Use Maintenance and Care of the Devices ...
Страница 37: ...85051B 4 1 4 Performance Verification ...
Страница 40: ...4 4 85051B Performance Verification Recertification ...
Страница 41: ...85051B 5 1 5 Troubleshooting ...
Страница 43: ...85051B 5 3 Troubleshooting Troubleshooting Process Figure 5 1 Troubleshooting Flowchart ...
Страница 46: ...5 6 85051B Troubleshooting Contacting Agilent ...
Страница 47: ...85051B 6 1 6 Replaceable Parts ...
Страница 50: ...6 4 85051B Replaceable Parts Introduction Figure 6 1 Replaceable Parts ...