2-2
85032F
Specifications
Environmental Requirements
Environmental Requirements
NOTE
Samples of this product have been type-tested in accordance with the Agilent
Environmental Test Manual and verified to be robust against the
environmental stresses of storage, transportation and end-use; those stresses
include but are not limited to temperature, humidity, shock, vibration,
altitude and power-line conditions. Test methods are aligned with IEC
60068-2 and levels are similar to MIL-PRF-28800F Class 3.
Temperature—What to Watch Out For
Changes in temperature can affect electrical characteristics. Therefore, the operating
temperature is a critical factor in performance. During a measurement calibration, the
temperature of the calibration devices must be stable and within the range shown in
Table 2-1.
IMPORTANT
Avoid unnecessary handling of the devices during calibration because your
fingers are a heat source.
Table 2-1
Environmental Requirements
Parameter Limits
Temperature
Operating
a
a. The temperature range over which the calibration standards maintain conformance to their
specifications.
+20
°
C to +26
°
C
Storage
−
40
°
C to +75
°
C
Error-corrected range
b
b. The allowable network analyzer ambient temperature drift during measurement calibration
and during measurements when the network analyzer error correction is turned on. Also, the
range over which the network analyzer maintains its specified performance while correction
is turned on.
±
1
°
C of measurement calibration temperature
Altitude
Operating
< 4,500 meters (
≈
15,000 feet)
Storage
< 4,500 meters (
≈
15,000 feet)
Relative humidity
Type tested, 0% to 95% at 40
°
C, non-condensing
Содержание 85032F
Страница 4: ...iv 85032F ...
Страница 7: ...1 1 1 General Information ...
Страница 12: ...1 6 85032F General Information Preventive Maintenance ...
Страница 13: ...2 1 2 Specifications ...
Страница 18: ...2 6 85032F Specifications Electrical Specifications ...
Страница 19: ...3 1 3 Use Maintenance and Care of the Devices ...
Страница 28: ...3 10 85032F Use Maintenance and Care of the Devices Gaging Connectors Figure 3 3 Gaging Type N Connectors ...
Страница 33: ...4 1 4 Performance Verification ...
Страница 36: ...4 4 85032F Performance Verification Recertification ...
Страница 37: ...5 1 5 Troubleshooting ...
Страница 41: ...6 1 6 Replaceable Parts ...
Страница 45: ...A 1 A Standard Definitions ...