Getting Started
Step 2. Install the Plug-in Module
Step 2. Install the Plug-in Module
You do
not
need to turn off the mainframe to install or remove the plug-in
modules.
The plug-in module can be installed in slots 1 and 2 or 3 and 4 on the
Agilent 83480A or Agilent 54750A mainframe. The plug-in module will
not
function if it is installed in slots 2 and 3.
To make sure the analyzer meets all of the published specifications, there
must be a good ground connection from the plug-in module to the mainframe.
The RF connectors on the rear of the plug-in module are spring-loaded, so fin-
ger-tighten the knurled screw on the front panel of the plug-in module to
make sure the plug-in is securely seated in the mainframe.
C A U T I O N
Do not use non-Agilent Technologies extender cables to operate the plug-in
module outside of the mainframe. The plug-in module can be damaged by
improper grounding when using extender cables.
Note
If you wish to use the Agilent 83486A optical plug-in module in an Agilent 54750A digi-
tizing oscilloscope, a firmware upgrade must first be installed. Order the Agilent 83480K
communications firmware kit and follow the installation instructions.
Note
The Agilent 83486A requires firmware revision A.06.0 or higher.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание 83486A
Страница 9: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 23: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 36: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 85: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 89: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...