2-13
Making Measurements
Cleaning Connections for Accurate Measurements
Figure 2-2. Universal adapters to Diamond HMS-10.
The HMS-10 encases the fiber within a soft nickel silver (Cu/Ni/Zn) center
which is surrounded by a tough tungsten carbide casing, as shown in
.
Figure 2-3. Cross-section of the Diamond HMS-10 connector.
The nickel silver allows an active centering process that permits the glass fiber
to be moved to the desired position. This process first stakes the soft nickel
silver to fix the fiber in a near-center location, then uses a post-active staking
to shift the fiber into the desired position within 0.2
µ
m. This process, plus the
keyed axis, allows very precise core-to-core alignments. This connector is
found on most Agilent Technologies lightwave instruments.
Содержание 83437A
Страница 1: ...Agilent 83438A Erbium ASE Source User s Guide ...
Страница 5: ...v The Agilent 83438A At a Glance Rear view of instrument ...
Страница 8: ......
Страница 10: ......
Страница 24: ...2 4 Making Measurements Performing Stimulus Response Measurements ...
Страница 41: ...3 Specifications 3 3 Regulatory Information 3 6 Specifications and Regulatory Information ...
Страница 47: ...3 7 Specifications and Regulatory Information Regulatory Information Declaration of Conformity I ll ...
Страница 48: ......
Страница 54: ......
Страница 61: ...5 7 Servicing General Information ...
Страница 63: ...5 9 Servicing General Information ...
Страница 79: ...5 25 Servicing Adjustment Procedure ...
Страница 85: ...5 31 Servicing Replacing Instrument Assemblies Location of resistors R2 R8 and R9 ...
Страница 92: ...5 38 Servicing Replaceable Parts ...
Страница 94: ...5 40 Servicing Replaceable Parts ...
Страница 96: ...5 42 Servicing Replaceable Parts ...
Страница 98: ...5 44 Servicing Replaceable Parts ...
Страница 100: ...5 46 Servicing Replaceable Parts ...
Страница 106: ......