85
Programming Reference
SCPI Instrument Command List
Examples
To set up a triggered burst of 16 Single Pulses at Out1, each burst
triggered by a positive edge at the EXT INPUT:
To set up gated pulses (Single Pulses at Out1), gated by a positive level at
the EXT INPUT:
Command
:TRIG:IMP
Long
:TRIGger:IMPedance
Form
Set & Query
Parameter
Numeric
Parameter Suffix
OHM
with engineering prefixes, e.g.:
MOHM
is Megaohms.
*RST value
50
Ω
Specified Limits
50
Ω
or 10 k
Ω
Description
Use this command to program the input impedance of the CLK-IN
connector. Note that only two settings are available. If you try to program
any other value, it will be rounded to one of the specified values.
Example
To set the input impedance and the threshold of the CLK-IN connector:
:ARM:SOUR EXT1
:ARM:SENS EDGE
:ARM:SLOP POS
:TRIG:COUN 16
:TRIG:SOUR INT1
:DIG:PATT OFF
:PULS:DOUB OFF
Set arming from EXT INPUT
Set arming on edges
Set arming on positive edges
Burst length 16
Pulse period trigger from internal osc.
Disable pattern operating mode
Ensure single pulses at OUTPUT
:ARM:SOUR EXT1
:ARM:SENS LEV
:ARM:SLOP POS
:TRIG:COUN 1
:TRIG:SOUR INT1
:DIG:PATT OFF
:PULS:DOUB OFF
Set arming from EXT INPUT
Set arming on levels
Set arming on positive level 1 pulse period
Single pulse output mode
Pulse period trigger from internal osc.
Disable pattern data
Ensure single pulses at OUTPUT
:TRIG:IMP 50OHM
:TRIG:LEV 2.5V
:TRIG:SOUR EXT2
Set CLK-IN impedance to 50
Ω
Set CLK-IN threshold to 2.5V
Pulse period trigger from CLK-IN
Содержание 81101A
Страница 1: ...S S1 Agilent 81101A 50 MHz Pulse Generator Reference Guide ...
Страница 88: ...88 Programming Reference SCPI Instrument Command List ...
Страница 112: ...112 Specifications Pulse Parameter Definitions ...
Страница 116: ...116 Index ...
Страница 118: ...Copyright Agilent Technologies 1998 2000 Edition E0300 Printed in Germany 81101 91021 ...