![Agilent Technologies 81101A Скачать руководство пользователя страница 65](http://html.mh-extra.com/html/agilent-technologies/81101a/81101a_reference-manual_2868455065.webp)
65
Programming Reference
SCPI Instrument Command List
Command
:PULS:TRAN[1]:UNIT
Long
[:SOURce]:PULSe:TRANsition[1]:UNIT
Form
Set & Query
Parameter
S | SEC | PCT
*RST value
S
Description
Use this command to set the default units for the pulse transition times.
The default unit is used when the parameter is programmed to a value
without a unit suffix.
Command
:PULS:TRAN[1]
Long
[:SOURce]:PULSe:TRANsition[1][:LEADing]
Form
Set & Query
Parameter
Numeric
Parameter suffix
S
with engineering prefixes, or
PCT
*RST value
5 ns
Specified limits
5 ns to 200 ms
Parameter
coupling
By default: Trailing edge = Leading edge with
:PULS:TRAN:TRA:AUTO
ON
.
Use
:PULS:TRAN:TRA:AUTO OFF
to enable independent programming
of the trailing edge within a 1:20 ratio for the ranges.
Description
Use this command to set/read the transition time of the pulse leading
edge. Note that the leading and trailing edges of the pulse have to fit
within the defined pulse width.
Example
To set leading and trailing edges independently:
:PULS:TRAN1 6NS
:PULS:TRAN:TRA:AUTO OFF
:PULS:TRAN:TRA 15 NS
Set OUTPUT leading edge to 6 ns
Enable independent setting of trailing
edge
Set OUTPUT trailing edge to 15 ns
Содержание 81101A
Страница 1: ...S S1 Agilent 81101A 50 MHz Pulse Generator Reference Guide ...
Страница 88: ...88 Programming Reference SCPI Instrument Command List ...
Страница 112: ...112 Specifications Pulse Parameter Definitions ...
Страница 116: ...116 Index ...
Страница 118: ...Copyright Agilent Technologies 1998 2000 Edition E0300 Printed in Germany 81101 91021 ...