![Agilent Technologies 81101A Скачать руководство пользователя страница 68](http://html.mh-extra.com/html/agilent-technologies/81101a/81101a_reference-manual_2868455068.webp)
68
Programming Reference
SCPI Instrument Command List
Command
:PULS:WIDT[1]
Long
[:SOURce]:PULSe:WIDTh[1]
Form
Set & Query
Parameter
Numeric
Parameter suffix
S
with engineering prefixes
*RST value
100 ns
Specified limits
10 ns to 999.5 s (max. period –10 ns)
Description
Use this command to program the width of the pulse signal. If you want
to set width as duty cycle use
[:SOURce]:PULSe:DCYCle[1]
.
If you want the pulse width to remain constant when the pulse period is
varied (rather than the duty cycle) use
[:SOURce]:PULSe:HOLD[1] WIDTh
.
Example
To set and hold the pulse width:
Command
:ROSC:SOUR
Long
[:SOURce]:ROSCillator:SOURce
Form
Set & Query
Parameter
INTernal | EXTernal
*RST value
INT
Description
Use this command to set/read the reference source for the PLL. If you
select the external reference (CLK-IN connector), you can choose to use
a 5 MHz or 10 MHz reference signal using
:ROSC:EXT:FREQ
.
:PULS:WIDT 50NS
:PULS:HOLD WIDTH
Set OUTPUT pulse width to 50 ns
Hold pulse width constant with varying period
INTernal
Lock the PLL to its internal reference
EXTernal
Lock the PLL to a reference signal at the
CLK-IN connector. The external refer-
ence signal can be 5 or 10 MHz.
Содержание 81101A
Страница 1: ...S S1 Agilent 81101A 50 MHz Pulse Generator Reference Guide ...
Страница 88: ...88 Programming Reference SCPI Instrument Command List ...
Страница 112: ...112 Specifications Pulse Parameter Definitions ...
Страница 116: ...116 Index ...