Troubleshooting
Tab. 2
SYMPTOM POSSIBLE
CAUSE
FIXING
3.
With the switch in ON
position, no LED are ON.
The battery is exhausted.
Change the battery.
4.
With the switch in ON
position, the LED L1 is
blinking and L2 is OFF.
The current that supplies
the pump is < 50 nA.
Discharge is not activated
and the pump does not
work.
Increase the pressure inside
the pump (ex. by heating the
pump), discharge is
activated and L1 goes OFF
and L2 starts blinking.
Ion Pump Battery Power Supply / 87-900-111-01 (E)
59/