2-21
Application Tutorials
Tutorial 6: Test to Industry Standards
Stop on and Display Trace Errors
The eye-diagram analyzer has the ability to stop data acquisition when a mask
violation occurs. The number of traces or errors that stop this data acquisition
can be specified. In addition, if you are in eyeline mode, you can separately
display the traces that have caused an error.
8
Set the bit interval and delay by pressing:
Setup
,
BIT INTVL
and enter
1
,
DELAY
and enter
–3
The mask will be offset to the right side of the display.
9
Set the analyzer so testing will stop after two errors have occurred by pressing:
#Errors ON OFF
ON and enter
2
10
If #Traces is set to on, the eye-diagram analyzer stops sweeping when either
the error or trace limit is reached. Turn the number of traces function off by
pressing:
#Traces ON OFF
OFF
11
Turn error tracing on by pressing:
err trc ON OFF
ON
Any trace which violates the mask shows on the lower-half of the screen.
12
Reset and start the trace and error counters by pressing:
Содержание 71501D
Страница 1: ...Agilent 71501D Eye Diagram Analysis User s Guide...
Страница 7: ...1 Getting Started...
Страница 34: ...Getting Started Steps for Setting Up Eye Diagram Analysis...
Страница 35: ...2 Application Tutorials...
Страница 43: ...2 9 Application Tutorials Tutorial 2 Measure in Optical Power Units...
Страница 48: ...Application Tutorials Tutorial 4 Measure Laser Turn on Delay Laser Overshoot and Turn On Delay...
Страница 80: ...Application Tutorials Tutorial 11 Create a Horizontal Histogram...
Страница 81: ...3 Eye Diagram Analyzer Reference...
Страница 93: ...3 13 Eye Diagram Analyzer Reference Eye Diagram Menu Maps Traces Measure Markers Masks and Mass Storage Menus...
Страница 113: ...3 33 Eye Diagram Analyzer Reference Applying Mask Testing The Mask Setup Submenu...
Страница 151: ...4 Programming Commands...
Страница 163: ...5 Specifications and Characteristics...
Страница 170: ...Specifications and Characteristics Declaration of Conformity Declaration of Conformity...