2-5
Application Tutorials
Tutorial 1: Measure Eye-Parameters
Example of a Large Overshoot Resulting from Laser Bias
Perform an Offset Calibration
This calibration procedure removes any offset that may be present in the opti-
cal-to-electrical converter. This is sometimes referred to as the "dark" level.
The offset calibration ensures accurate measurements of the laser’s one and
zero levels.
4
Turn the laser off. (If you are measuring the pattern generator directly,
disconnect the input signal at channel 1.)
5
Perform the calibration by pressing:
Trg,Cal
,
OFFSET CAL
,
CONT
The calibration takes about a minute to execute. When the calibration is fin-
ished,
DC NULL: done
is displayed.
6
Turn the laser on. (If you are measuring the pattern generator directly,
reconnect the input signal at channel 1.)
Measure Signal Parameters
7
To include the rise time and fall time measurement in the displayed results,
press:
Measure
,
r/f tim ON OFF
ON
Содержание 71501D
Страница 1: ...Agilent 71501D Eye Diagram Analysis User s Guide...
Страница 7: ...1 Getting Started...
Страница 34: ...Getting Started Steps for Setting Up Eye Diagram Analysis...
Страница 35: ...2 Application Tutorials...
Страница 43: ...2 9 Application Tutorials Tutorial 2 Measure in Optical Power Units...
Страница 48: ...Application Tutorials Tutorial 4 Measure Laser Turn on Delay Laser Overshoot and Turn On Delay...
Страница 80: ...Application Tutorials Tutorial 11 Create a Horizontal Histogram...
Страница 81: ...3 Eye Diagram Analyzer Reference...
Страница 93: ...3 13 Eye Diagram Analyzer Reference Eye Diagram Menu Maps Traces Measure Markers Masks and Mass Storage Menus...
Страница 113: ...3 33 Eye Diagram Analyzer Reference Applying Mask Testing The Mask Setup Submenu...
Страница 151: ...4 Programming Commands...
Страница 163: ...5 Specifications and Characteristics...
Страница 170: ...Specifications and Characteristics Declaration of Conformity Declaration of Conformity...