Maintenance and Troubleshooting
Agilent 5800 and 5900 ICP-OES User's Guide
63
2
Check for carry-over after refitting the torch to the instrument to determine if
the cleaning procedure has been sufficient. If carry-over is found, repeat the
cleaning process.
3
Replace the torch when the outer surface of the quartz outer tube is rough to
the touch (which indicates signs of wear), or if there are any cracks visible.
NOTE
Long term exposure to acid during cleaning may lead to discoloration of the plastic
base. This change is cosmetic only and should not impact performance if the torch
is clean and the results of the other torch checks are satisfactory.
TIP
Store the torch in the original box or a plastic bag when not in use, or alternately,
use the Agilent ICP-OES Easy-fit torch storage rack (P/N G8010-67000).
Troubleshooting
For troubleshooting information, please see the ICP Expert Help and Learning
Center:
1
Double-click the ICP Expert Help and Learning Center on the computer
desktop.
2
Click on, or search for Troubleshooting to view instructions on how to
troubleshoot.
Spare Parts
For spare parts and consumables ordering information, refer to the Agilent
Technologies website:
www.agilent.com
To replace the items listed below, you must use Agilent-manufactured parts, which
can be ordered online from the Agilent website or through your local sales
representative.
Содержание 5800 ICP-OES
Страница 1: ...Agilent 5800 and 5900 ICP OES User s Guide ...
Страница 6: ...Contents 6 Agilent 5800 and 5900 ICP OES User s Guide This page is intentionally left blank ...
Страница 38: ...Installation 38 Agilent 5800 and 5900 ICP OES User s Guide This page is intentionally left ...
Страница 50: ...Operation 50 Agilent 5800 and 5900 ICP OES User s Guide This page is intenionally blank ...
Страница 65: ......