Operation
Agilent 5800 and 5900 ICP-OES User's Guide
41
13
Press the Power on/off button on the front of the instrument. The power
on/off LED will display green when it is on. The ICP-OES is now in standby
state, which continually purges and thermostatically controls the
polychromator.
NOTE
Both the mains power switch on the left side of the instrument and front panel
power button must be on for the instrument to work.
14
Double-click the ICP Expert desktop icon to start the ICP Expert software. The
Start page will appear.
Running the instrument from standby
During standby state (both power switches are ‘On’), the polychromator
temperature is regulated and polychromator gas purge is running to ensure the
instrument is ready for analysis within 20 minutes after lighting the plasma.
To begin using the instrument from standby state:
1
Turn on the laboratory exhaust system.
2
Check that the torch is clean and in good condition and installed with the torch
handle fully closed.
3
Check that all tubing on the spray chamber, nebulizer and peristaltic pump is
correctly connected.
4
Check that the torch compartment door is fully closed.
5
Switch on the monitor and printer (if they are off).
6
Switch on the water chiller (if it is off).
7
If you have accessories fitted, switch them on.
Preparing for Analysis
To prepare for analysis:
1
Click the Plasma button in the ICP Expert software. Alternatively, press F5 or
choose Plasma on from the arrow under the Plasma button.
Содержание 5800 ICP-OES
Страница 1: ...Agilent 5800 and 5900 ICP OES User s Guide ...
Страница 6: ...Contents 6 Agilent 5800 and 5900 ICP OES User s Guide This page is intentionally left blank ...
Страница 38: ...Installation 38 Agilent 5800 and 5900 ICP OES User s Guide This page is intentionally left ...
Страница 50: ...Operation 50 Agilent 5800 and 5900 ICP OES User s Guide This page is intenionally blank ...
Страница 65: ......