Chapter 4: Calibrating and Adjusting
To check the power supply
4–5
4
Connect the common lead of the voltmeter to the GND test point.
5
Connect the positive lead of the voltmeter to the +5.1 V test point.
6
Verify that the +5.1 supply voltage is within limits as shown in the following table:
Table 4-1
Power Supply Voltage Limits
7
Repeat steps 4 through 6 for each of the other supply voltages.
If any supply voltage is not within specifications, see chapter 5, “Troubleshooting.”
Supply Voltage Specification
Limits
+5.1 V
±
0.1 V
+5.0 V to +5.2 V
-5.2 V
±
0.1 V
-5.1 V to -5.3 V
+12.2 V
±
0.3 V
+11.9 V to +12.5 V
-12.2 V
±
0.3 V
-11.9 V to -12.5 V
+15 V bias
+14 V to +16 V
Содержание 54835A
Страница 8: ...Contents 4 ...
Страница 23: ...Chapter 1 General Information Agilent Technologies 54835A 45A 46A general characteristics 1 15 ...
Страница 64: ...3 26 ...
Страница 78: ...4 12 ...
Страница 128: ...Chapter 5 Troubleshooting To troubleshoot the acquisition system 5 50 Acquisition Trouble Isolation 1 of 4 ...
Страница 129: ...Chapter 5 Troubleshooting To troubleshoot the acquisition system 5 51 Acquisition Trouble Isolation 2 of 4 ...
Страница 130: ...Chapter 5 Troubleshooting To troubleshoot the acquisition system 5 52 Acquisition Trouble Isolation 3 of 4 ...
Страница 138: ...5 60 ...
Страница 182: ...Chapter 7 Replaceable Parts Exploded Views 7 10 120 Mbyte Floppy Drive Configuration for PC Motherboard Configuration 4 ...
Страница 186: ...Chapter 7 Replaceable Parts Exploded Views 7 14 Attenuator Assembly ...
Страница 194: ...7 22 ...
Страница 196: ...Chapter 8 Theory of Operation 8 2 Instrument Block Diagram Shown for Original Configuration w 1 44 MByte Floppy Drive ...
Страница 200: ...Chapter 8 Theory of Operation Block Level Theory 8 6 Acquisition Block Diagram CHANNEL 1 2 TIME BASE TRIGGER ...
Страница 204: ...8 10 ...