Chapter 3: Testing Performance
To test voltage measurement accuracy
3–8
Figure 3-4
Voltage Measurement Accuracy Equipment Setup
2
Press Default Setup to set the oscilloscope to default conditions.
3
Set all channels to dc using the Coupling key and to 1 M
Ω
input impedance using the
Input key.
4
Using the mouse, enable the graphical interface.
5
Select Acquisition from the Setup menu.
6
Select Equivalent Time sampling mode. Enable Averaging. Set the # of Averages to 32,
either using the spin box or the numeric keypad dialog. Click Close.
See figure 3-5.
Figure 3-5
Acquisition Setup for Voltage Accuracy Measurement
7
Use the following table for steps 8 through 15.
Select Equivalent Time
sampling mode
Enable averaging
Set the number of
averages to 32
Leave Memory Depth
on Automatic
Содержание 54835A
Страница 8: ...Contents 4 ...
Страница 23: ...Chapter 1 General Information Agilent Technologies 54835A 45A 46A general characteristics 1 15 ...
Страница 64: ...3 26 ...
Страница 78: ...4 12 ...
Страница 128: ...Chapter 5 Troubleshooting To troubleshoot the acquisition system 5 50 Acquisition Trouble Isolation 1 of 4 ...
Страница 129: ...Chapter 5 Troubleshooting To troubleshoot the acquisition system 5 51 Acquisition Trouble Isolation 2 of 4 ...
Страница 130: ...Chapter 5 Troubleshooting To troubleshoot the acquisition system 5 52 Acquisition Trouble Isolation 3 of 4 ...
Страница 138: ...5 60 ...
Страница 182: ...Chapter 7 Replaceable Parts Exploded Views 7 10 120 Mbyte Floppy Drive Configuration for PC Motherboard Configuration 4 ...
Страница 186: ...Chapter 7 Replaceable Parts Exploded Views 7 14 Attenuator Assembly ...
Страница 194: ...7 22 ...
Страница 196: ...Chapter 8 Theory of Operation 8 2 Instrument Block Diagram Shown for Original Configuration w 1 44 MByte Floppy Drive ...
Страница 200: ...Chapter 8 Theory of Operation Block Level Theory 8 6 Acquisition Block Diagram CHANNEL 1 2 TIME BASE TRIGGER ...
Страница 204: ...8 10 ...