Chapter 6: Replacing Assemblies
To remove and replace the AutoProbe assembly
6–6
To remove and replace the AutoProbe assembly
Use this procedure to remove and replace the AutoProbe assembly. When necessary, refer to
other removal procedures.
1
Disconnect the power cable and remove the cover.
2
Remove the AutoProbe connector assembly, the subpanel, and the probe connector
assembly that fits around the front-panel BNC connectors, by doing the following:
a
Disconnect the mylar ribbon cable W13 from the Probe Power and Control Board, A11.
See “To disconnect and connect Mylar flex cables” in this chapter.
Figure 6-4
Disconnecting W13
b
Locate the access hole in the front-panel assembly below and almost between channel
two and channel three attenuators.
Figure 6-5
Access Hole
Содержание 54835A
Страница 8: ...Contents 4 ...
Страница 23: ...Chapter 1 General Information Agilent Technologies 54835A 45A 46A general characteristics 1 15 ...
Страница 64: ...3 26 ...
Страница 78: ...4 12 ...
Страница 128: ...Chapter 5 Troubleshooting To troubleshoot the acquisition system 5 50 Acquisition Trouble Isolation 1 of 4 ...
Страница 129: ...Chapter 5 Troubleshooting To troubleshoot the acquisition system 5 51 Acquisition Trouble Isolation 2 of 4 ...
Страница 130: ...Chapter 5 Troubleshooting To troubleshoot the acquisition system 5 52 Acquisition Trouble Isolation 3 of 4 ...
Страница 138: ...5 60 ...
Страница 182: ...Chapter 7 Replaceable Parts Exploded Views 7 10 120 Mbyte Floppy Drive Configuration for PC Motherboard Configuration 4 ...
Страница 186: ...Chapter 7 Replaceable Parts Exploded Views 7 14 Attenuator Assembly ...
Страница 194: ...7 22 ...
Страница 196: ...Chapter 8 Theory of Operation 8 2 Instrument Block Diagram Shown for Original Configuration w 1 44 MByte Floppy Drive ...
Страница 200: ...Chapter 8 Theory of Operation Block Level Theory 8 6 Acquisition Block Diagram CHANNEL 1 2 TIME BASE TRIGGER ...
Страница 204: ...8 10 ...