4- 5
Calibrating and Adjusting
To adjust the power supply
2
Locate the power supply voltages at L200, L201, L202, and L203 on the
system board.
The power supply voltages are not labeled on the system board. See
Figure 4-1 for the locations.
Figure 4-1
Low Voltage Power Supply Voltages (on the bottom of the oscilloscope)
3
Make sure that the voltage measurements are within the tolerances
listed in Table 4-2.
Table 4-2
Power Supply Voltage Tolerances
E
-
-
-
9
/
9
9
9
-
-
/
/
/
Supply Voltage
Tolerance
+5.1 V
±250 mV (4.85 V to 5.35 V)
-5.2 V
±156 mV (-5.04 V to -5.36 V)
+15.75 V
+1.260 V, -787 mV (+14.963 V to +17.010 V)
+3.4 V
±100 mV (+3.30 V to +3.50 V)
+8.2 V
±82 mV (+8.12 V to +8.28 V)
measure at Printer Power connector on rear panel
of instrument
service.book Page 5 Wednesday, August 28, 2002 3:25 PM
Содержание 54641A
Страница 4: ...iv service book Page iv Wednesday August 28 2002 3 25 PM ...
Страница 7: ...1 General Information ...
Страница 24: ...1 18 ...
Страница 25: ...2 Preparing the Oscilloscope for Use service book Page 1 Wednesday August 28 2002 3 25 PM ...
Страница 45: ...3 Testing Performance service book Page 1 Wednesday August 28 2002 3 25 PM ...
Страница 66: ...3 22 service book Page 22 Wednesday August 28 2002 3 25 PM ...
Страница 67: ...4 Calibrating and Adjusting service book Page 1 Wednesday August 28 2002 3 25 PM ...
Страница 76: ...4 10 service book Page 10 Wednesday August 28 2002 3 25 PM ...
Страница 77: ...5 Troubleshooting service book Page 1 Wednesday August 28 2002 3 25 PM ...
Страница 90: ...5 14 service book Page 14 Wednesday August 28 2002 3 25 PM ...
Страница 91: ...6 Replacing Assemblies service book Page 1 Wednesday August 28 2002 3 25 PM ...
Страница 107: ...7 Replaceable Parts service book Page 1 Wednesday August 28 2002 3 25 PM ...
Страница 114: ...7 8 service book Page 8 Wednesday August 28 2002 3 25 PM ...