Installing the NPD/ECD Chemical Filters Accessory 19199N
Preparing the Instrument
4
10. Disconnect the outlet fitting plate from the top of the filters. Remove the
flow lines from the filters. Route these lines to the detector flow
manifold.
WARNING
Hydrogen (H
2
) is flammable and is an explosion hazard when confined in an
enclosed space such as the oven. In any application using hydrogen, turn off
the supply at its source before working on the instrument.
Air, NPD (5890 only)
Anode purge, ECD
Hydrogen, NPD (5890 only)
Makeup gas, ECD
Outlet fitting plate