Installing the NPD/ECD Chemical Filters
Accessory 19199N
Preparing the Instrument
WARNING
Hazardous voltages are present in the instrument whenever the power cord
is connected. Avoid a potentially dangerous shock hazard by disconnecting
the power cord before working on the instrument.
1.
Set the main power line switch to the off position.
2.
Disconnect the power cable from its receptacle.
3.
Allow time for the oven and heated zones to cool.
4.
When the heated zones are cool, turn off all gas supplies at the source.
5.
If an autosampler is installed on the instrument, it will be necessary to
remove it and its mounting bracket to allow removal of the left side
cover.
Red O visible
ON
OFF