Read this section carefully before you attempt to reinstall the operating system
from the CD-ROM using this procedure.
You will need to connect a keyboard to your 16702B in order to execute disaster
recovery procedures.
Everything on the hard drive will be
overwritten, including user configuration, data files, and license passwords.
To save your system's license information, as well as other system settings,
refer to “Saving System Settings” and “Reloading System Settings" in this
section.
A batch process is used to autoload the software and then reboot the instrument. The batch process waits
for only a short timeout period for user interaction to abort the process. Otherwise, the hard drive will be
initialized, the operating system will be uploaded, and the instrument will reboot.
The reinstallation process takes approximately one hour depending on the speed of the attached CD-ROM.
If required, follow the steps in this book to setup the instrument and CD-ROM drive.
Insert the CD-ROM containing the instrument operating software into the CD-ROM
drive. Allow a couple of moments for the media to settle after inserting the media.
If required, follow the steps in "Saving System Settings” and “Reloading System
Settings" to create a backup file of your system settings and license passwords.
If the LAN cable is connected, disconnect it from the instrument. If needed, turn on the
system and initiate the monitor selection mode. (
Otherwise,
proceed to step 4.
See the section in this book.)
Turn on the instrument and repeatedly press the [ESC] key on the keyboard to terminate
the boot process. When the boot process is terminated, a prompt will be displayed.
Press:
Type:
The instrument will search for all viable boot devices on the bus, including the CD-ROM
drive. The display will then show the boot devices:
Main Menu: Enter command >
<Enter>
SEA <Enter>
Path Number
Device Path
Device Type
------------
------------
------------------
P0
SESCSI.6.0
IBM
DNES-309170W
P1
SESCSI.1.0
PLEXTOR CD-ROM PX-40TS
Reinstalling the Operating System.