![Agilent Technologies 16700B Скачать руководство пользователя страница 28](http://html1.mh-extra.com/html/agilent-technologies/16700b/16700b_installation-manual_2867159028.webp)
16allp25
You can connect the logic analyzer cable directly to a 40-pin connector, but you must install the
proper RC network directly onto the target system board. Agilent Technologies recommends
three types of
which are described in detail in the Application Note:
RC networks
Probing
Solutions for Agilent Logic Analysis Systems.
For more information, contact your Agilent Technologies Sales office and ask for the
Application Note:
Probing Solutions for Agilent Logic Analysis Systems.
(Or
download from the web at: http://www.agilent.com/find/go/LA-AppNotes/ )
28
Connecting Probes to a Target System Directly
Target Connector Pinout
(Top View)
+5V
CLK1
N/C
D15
D14
D13
D12
D11
D10
D9
D8
D7
D6
D5
D4
D3
D2
D1
D0
+5V
1
3
5
7
9
11
13
15
17
19
21
23
25
27
29
31
33
35
37
39
2 Power GND
4 Signal GND
6 Signal GND
8 Signal GND
10 Signal GND
12 Signal GND
14 Signal GND
16 Signal GND
18 Signal GND
20 Signal GND
22 Signal GND
24 Signal GND
26 Signal GND
28 Signal GND
30 Signal GND
32 Signal GND
34 Signal GND
36 Signal GND
38 Signal GND
40 Power GND
Do not exceed 0.33 amps per cable, or the cable will be damaged. The cable ground lines
are chasis (earth) grounds and not "floating" grounds. All the lines are woven into a flat
ribbon that is 4.5 feet long.
CAUTION
Logic Analyzer
Pod
Suggested On Board RC Network
Equivalent Load
8.2pF
250
ohm
370
ohm
90.9k
ohm
Signal
Signal
To Logic
Analyzer
Pod
Ground
Ground
7.4pF
0.3pF
100k
ohm
Includes on board RC network and logic analyzer
Probing
for 16700B/ 16702B/ Measurement Modules