Socket Modules for Small Signal Devices
The six socket modules for small signal devices have FORCE terminals only. Up
to 3 SMU channels can be connected to DUT by Kelvin Connection. As shown
in the example in the following figure, use miniature banana to pin plug
connection cables to connect both FORCE and SENSE terminals of the 16442A
to the FORCE terminal on the socket module as follows: insert one pin plug into
the FORCE terminal of the socket module, then insert the other pin plug into the
tail hole of the inserted pin plug.
To connect sense terminals as near as possible to the DUT, connect the SENSE
terminals of the 16442A directly to the DUT with the miniature banana to
miniature clip connection cables as shown in following figure.
Содержание 16442A