HPLC-Chip Cube Interface User’s Guide
69
Operation
5
Spray Ignition
Conditions to avoid
Figure 27
Spray angled into the MS entrance
To correct the spray direction, do the following:
•
Lower the capillary voltage.
•
Turn the thumb wheel on the left bottom side of the Chip Cube. Move the
HPLC-Chip tip further towards the counter electrode until the spray looks
like
HPLC-Chip tip
counter electrode
end cap
nano spray
HV-capillary
C A U T I O N
The spray must not be angled directly into the MS entrance, this will contaminate the
desolvation assembly of the MS and cause high background.