User Manual SIM Sample Cooler_06/2018
(9) Equilibration Time:
the SIM Cooler requires the temperature to
be within a specifiable tolerance of the
setpoint for this time to get ready.
Switch Off after Sequence:
cooling of the samples automatically ends
when the sequence is finished.
Cooler Controlled by method:
you can specify methods in OpenLAB CDS
which control the temperature setpoint, the
readiness feature and the temperature
tolerance. If the box is unchecked, the
cooler can only be switched on and off using
the status dashboard - the temperature
setpoint can only be changed using a
programmed preset or a physical panel on
the device.