Agilent 1200 Series Autosampler Reference Manual
4
Repairing the Autosampler
Simple Repairs
The procedures described in this section can be done with the autosampler in
place in the stack. You will do some of these procedures on a more frequent
basis.
Table 16
Simple Repair Procedures
Procedure
Typical Frequency
Time Required
Notes
Exchanging the needle
assembly
When needle shows indication of
damage or blockage
15 minutes
See
Exchanging the seat
assembly
When the seat shows indication of
damage or blockage
10 minutes
See
Exchanging the rotor
seal
After approximately 30000 to 40000
injections, or when the valve
performance shows indication of
leakage or wear
30 minutes
See
Exchanging the
metering seal
When autosampler reproducibility
indicates seal wear
30 minutes
See
Exchanging the gripper
arm
When the gripper arm is defective
10 minutes
See
Содержание 1200 series
Страница 1: ...A Agilent 1200 Series Standard and Preparative Autosamplers Reference Manual ...
Страница 12: ...12 Agilent 1200 Series Autosampler Reference Manual Contents ...
Страница 42: ...42 Agilent 1200 Series Autosampler Reference Manual 1 Installing the Autosampler ...
Страница 104: ...104 Agilent 1200 Series Autosampler Reference Manual 3 Troubleshooting and Test Functions ...
Страница 108: ...108 Agilent 1200 Series Autosampler Reference Manual 4 Repairing the Autosampler Figure 14 Using the ESD Strap ...
Страница 174: ...174 Agilent 1200 Series Autosampler Reference Manual 4 Repairing the Autosampler ...
Страница 198: ...198 Agilent 1200 Series Autosampler Reference Manual 5 Parts and Materials ...
Страница 232: ...232 Agilent 1200 Series Autosampler Reference Manual 7 Introduction to the Autosampler ...
Страница 277: ......