124
Agilent 1200 Series Autosampler Reference Manual
4
Repairing the Autosampler
Metering Seal and Plunger
Frequency
Poor injection-volume reproducibility
Leaking metering device
Tools required
1/4 inch wrench (supplied in accessory kit).
4 mm hex key (supplied in accessory kit).
3 mm hex key (supplied in accessory kit).
Parts required
Metering seal 5063-6589 (pack of 2) for 100
µl analytical head
Metering seal 0905-1294 (pack of 1) for 900
µl analytical head
Metering plunger 5063-6586 for 100
µl analytical head
Metering plunger 5062-8587 for 900
µl analytical head
(only if scratched or contaminated)
Preparations for this
Procedure
Select “Start” in the maintenance function “Change piston” (see
Remove the front cover.
1 Remove the two capillaries from the metering-head
assembly.
2 Remove the two fixing bolts, and pull the head assembly
away from the sampler. Notice that the closed side of the
metering head faces upwards.
CellFrame
CellFrame
Содержание 1200 series
Страница 1: ...A Agilent 1200 Series Standard and Preparative Autosamplers Reference Manual ...
Страница 12: ...12 Agilent 1200 Series Autosampler Reference Manual Contents ...
Страница 42: ...42 Agilent 1200 Series Autosampler Reference Manual 1 Installing the Autosampler ...
Страница 104: ...104 Agilent 1200 Series Autosampler Reference Manual 3 Troubleshooting and Test Functions ...
Страница 108: ...108 Agilent 1200 Series Autosampler Reference Manual 4 Repairing the Autosampler Figure 14 Using the ESD Strap ...
Страница 174: ...174 Agilent 1200 Series Autosampler Reference Manual 4 Repairing the Autosampler ...
Страница 198: ...198 Agilent 1200 Series Autosampler Reference Manual 5 Parts and Materials ...
Страница 232: ...232 Agilent 1200 Series Autosampler Reference Manual 7 Introduction to the Autosampler ...
Страница 277: ......