88
Agilent 1200 Series Autosampler Reference Manual
3
Troubleshooting and Test Functions
Change Piston
The change-piston function draws the piston away from the home position,
relieving the tension on the spring. In this position, the analytical head
assembly can be removed and reinstalled easily after maintenance.
User Interface
The commands for the ChemStation and Control Module (different names for
commands in the Control Module are shown in brackets) are:
Start (Change)
Draws the piston away from the home position, relieving the tension on the
spring.
End (Done)
Repositions the plunger at the home position.
Using the Change Seal Function
1
Select “
Start
” (“
Change
") to move the piston to the maintenance position.
2
Exchange the metering seal (see
3
Select “
End
” (“
Done
”) to move the piston back to the home position.
Содержание 1200 series
Страница 1: ...A Agilent 1200 Series Standard and Preparative Autosamplers Reference Manual ...
Страница 12: ...12 Agilent 1200 Series Autosampler Reference Manual Contents ...
Страница 42: ...42 Agilent 1200 Series Autosampler Reference Manual 1 Installing the Autosampler ...
Страница 104: ...104 Agilent 1200 Series Autosampler Reference Manual 3 Troubleshooting and Test Functions ...
Страница 108: ...108 Agilent 1200 Series Autosampler Reference Manual 4 Repairing the Autosampler Figure 14 Using the ESD Strap ...
Страница 174: ...174 Agilent 1200 Series Autosampler Reference Manual 4 Repairing the Autosampler ...
Страница 198: ...198 Agilent 1200 Series Autosampler Reference Manual 5 Parts and Materials ...
Страница 232: ...232 Agilent 1200 Series Autosampler Reference Manual 7 Introduction to the Autosampler ...
Страница 277: ......