Agilent 1200 Series Autosampler Reference Manual
111
Repairing the Autosampler
4
Needle Assembly
Frequency
When the needle is visibly damaged
When the needle is blocked
Tools required
¼
inch wrench (supplied in accessory kit)
2.5 mm Hex key (supplied in accessory kit)
A pair of pliers
Parts required
Needle assembly (G1313-87201) for G1313-87101 or G1313-87103 needle-seat
Needle assembly (G1329-80001) for G1329-87101 or G1329-87103 needle seat
Needle assembly (900 µl loop cap) (G1313-87202) for G1313-87101 needle
seat
Needle assembly (900 µl loop cap) (G2260-87201 ) for G2260-87101
needle-seat
Preparations for this
procedure
Select “Start” in the maintenance function “Change Needle” (see
When the needle is positioned approx.15 mm above the needle seat, remove
the front cover.
WA R N I N G
To avoid personal injury, keep fingers away from the needle area during autosampler
operation. Do not bend the safety flap away from its position, or attempt to insert or
remove a vial from the gripper when the gripper is positioned below the needle.
Содержание 1200 series
Страница 1: ...A Agilent 1200 Series Standard and Preparative Autosamplers Reference Manual ...
Страница 12: ...12 Agilent 1200 Series Autosampler Reference Manual Contents ...
Страница 42: ...42 Agilent 1200 Series Autosampler Reference Manual 1 Installing the Autosampler ...
Страница 104: ...104 Agilent 1200 Series Autosampler Reference Manual 3 Troubleshooting and Test Functions ...
Страница 108: ...108 Agilent 1200 Series Autosampler Reference Manual 4 Repairing the Autosampler Figure 14 Using the ESD Strap ...
Страница 174: ...174 Agilent 1200 Series Autosampler Reference Manual 4 Repairing the Autosampler ...
Страница 198: ...198 Agilent 1200 Series Autosampler Reference Manual 5 Parts and Materials ...
Страница 232: ...232 Agilent 1200 Series Autosampler Reference Manual 7 Introduction to the Autosampler ...
Страница 277: ......