120
Agilent 1200 Series Autosampler Reference Manual
4
Repairing the Autosampler
Rotor Seal
Frequency
Poor injection-volume reproducibility
Leaking injection valve
Tools required
1/4 inch wrench (supplied in accessory kit).
Hex key, 9/64 inch (supplied in accessory kit).
Parts required
Rotor seal 0100-1853 (Vespel) for G1329A
Rotor seal 0100-1849 (Tefzel) for G1329A
Rotor seal 0101-1268 (PEEK) for G2260A
Preparations for this
procedure
Remove front cover.
Remove the leak tubing (if necessary).
C A U T I O N
The stator face is held in place by the stator head. When removing the stator head,
ensure the stator face does not fall out of the valve.
1 Remove all capillary fittings from the injection-valve ports.
2 Loosen each fixing bolt two turns at a time. Remove the
bolts from the head.
Содержание 1200 series
Страница 1: ...A Agilent 1200 Series Standard and Preparative Autosamplers Reference Manual ...
Страница 12: ...12 Agilent 1200 Series Autosampler Reference Manual Contents ...
Страница 42: ...42 Agilent 1200 Series Autosampler Reference Manual 1 Installing the Autosampler ...
Страница 104: ...104 Agilent 1200 Series Autosampler Reference Manual 3 Troubleshooting and Test Functions ...
Страница 108: ...108 Agilent 1200 Series Autosampler Reference Manual 4 Repairing the Autosampler Figure 14 Using the ESD Strap ...
Страница 174: ...174 Agilent 1200 Series Autosampler Reference Manual 4 Repairing the Autosampler ...
Страница 198: ...198 Agilent 1200 Series Autosampler Reference Manual 5 Parts and Materials ...
Страница 232: ...232 Agilent 1200 Series Autosampler Reference Manual 7 Introduction to the Autosampler ...
Страница 277: ......