3. The full Q band display is shown in
Figure 3-12.
4. To complete the amplitude calibration process, the preselector must be adjusted at each
frequency of interest. Before making final amplitude measurements with the analyzer,
perform the following:
a. Place a marker on the signal of interest.
b. Press
SPAN
,
Span Zoom
, 10,
MHz
to zoom in on the signal.
c. Press
AMPLITUDE
,
Presel Center
.
The final amplitude measurement can now be read out with the marker. See
.
Figure 3-13.