58
Chapter 3
Operation
Using an 8566B Spectrum Analyzer
O
p
er
ati
o
n
Using an 8566B Spectrum Analyzer
Preliminary Operation
Compatibility Reference
Refer to
on page 27 for information relating to retrofit
requirements for 8566A and older 8566B Spectrum Analyzers.
Equipment Connection
Connect the equipment as shown in
on page 59 and use the following checklist
to verify the connections.
• 11974 LO IN to 11975A OUTPUT (SMA cable, 5061-5458)
• 11974 IF OUT to 8566B IF INPUT (SMA cable, 5061-5458)
• 11974 TUNE IN to 8566B SWEEP + TUNE OUT (BNC cable, 8120-1839; BNC f–f
adapter, 1250-0080; BNC cable, 8120-2582)
• 11974 POWER SUPPLY to 11974-60028 PRESELECTOR POWER
• 8566B 1ST LO OUTPUT to 11975A INPUT (SMA cable, 5061-5458)
CAUTION
When you are using an 11975A Amplifier with an 11974, you must set the
amplifier rear-panel ALC switch to ON before you connect the amplifier into
the test setup. If the ALC switch is left in the OFF position, the amplifier
output power is high enough to destroy the mixer diodes.
Set the 11975A output power to +16 dBm and verify that the amplifier's front-panel
unleveled indicator is not lit.
Содержание 11974 Series
Страница 5: ...4 ...
Страница 8: ...7 General Information 1 General Information ...
Страница 28: ...27 Installation 2 Installation ...
Страница 47: ...46 Chapter2 Installation Returning Equipment for Service Installation ...
Страница 48: ...47 Operation 3 Operation ...
Страница 57: ...56 Chapter3 Operation Operating Hints Operation Figure 3 4 Setup Configuration for Remote Location of an 11974 ...
Страница 97: ...96 Chapter3 Operation Using 8561A 62A 62B Spectrum Analyzers Operation ...
Страница 98: ...97 Performance Tests 4 Performance Tests ...
Страница 107: ...106 Chapter4 Performance Tests Performance Test Procedures Performance Tests Figure 4 1 Performance Tests Setup ...
Страница 148: ...147 Safety and Regulatory Information 5 Safety and Regulatory Information ...