18
1145A 2-Channel Active Probe
Bandwidth
Bandwidth
Specification
down less than 3dB, dc to 750 MHz
Equipment Required
Equipment
Critical Specification
Recommended
Model/Part
Signal Generator
50 MHz to 800 MHz
8663A
Power Meters (2) or
one Dual-Channel
50 MHz to 800 MHz
±3% accuracy
436A (2),
437A (2), or
438A (1)
(or equivalent)
Power Sensor (2)
50 MHz to 800 MHz, 300 mW
8482A
Power Splitter
Type-N, dc to 800 MHz,
≤
0.2 dB tracking
11667A
Power Supply
Power for probe under test
Oscilloscope or
1142A
Termination
SMA feedthrough, 50
Ω
0955-0736
Adapter*
Type N(m) to SMA(m), 50
Ω
1250-1994 (or
equivalent)
Adapter*
SMA(f)-to-probe
16517-27601
Adapter
BNC(m) to N(f), 50
Ω
1250-0077
Adapter
N(m-m), 50
Ω
1250-0778
*Part of calibration kit part number Agilent Technologies 01145-68701