16
1145A 2-Channel Active Probe
Input Resistance
Input Resistance
Specification
Equipment Required
1
Connect the DMM between the probe tip and the ground at the tip of
the probe.
2
Set up the DMM to measure resistance.
DC Gain Accuracy
Specification
Equipment Required
Equipment
Critical Specification
Recommended
Model/Part
Digital Multimeter
Resistance ±1%
34401A
Equipment
Critical Specification
Recommended
Model/Part
Digital Multimeter
Better than 0.1% accuracy at 1 KHz34401A
Power Supply
1 V ±1%
6114A (or equivalent)
Power Supply
Power for probe under test
Oscilloscope or
1142A
Termination
BNC feedthrough, 50
Ω
±0.5%
11048C
Adapter
BNC (f) to banana (m)
1251-2277