MONITOR WITH SPECIAL CARE THE PATIENT POSITION OR
ANYONE PRESENT, TO AVOID INJURY CAUSED BY UNIT
MOVEMENTS.
INTRAVENOUS TUBING, CATHETERS AND OTHER PATIENT
CONNECTED LINES SHOULD BE ROUTED AWAY FROM
MOVING EQUIPMENT.
Motion Controls are only enabled when the Switch-Key on the
Control Panel is in the “ON” position.
Always place the Unit in Parking position before turning the
Generator and Console off, even though lock controls will
remain enabled for 15 seconds after turning off both the
Generator and Console in order to place the Unit in Parking
position.