DX-D 100 Mobile X-ray Unit
User Manual
32
0188K EN 20201116
GUIDANCE AND MANUFACTURER’S DECLARATION - ELECTROMAGNETIC IMMUNITY
(IEC 60601-1-2:2014)
This X-ray System is intended for use in the electromagnetic environment specified below.
The customer or user of this X-ray System should assure that it is used in such an environment.
Immunity Test
IEC 60601-1-2:2014
Test Level
Compliance Level
Electromagnetic
environment - guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
¦
8 kV contact
¦
2 kV,
¦
kV,
¦
8 kV,
¦
15 kV air
¦
8 kV contact
¦
2 kV,
¦
kV,
¦
8 kV,
¦
15 kV air
Floors
should
be
wood,
concrete or ceramic tile. If floors
are covered with synthetic
material, the relative humidity
should be at least 30 %.
Electrical fast transient/burst
IEC 61000-4-4
¦
2 kV for power supply lines
¦
1 kV for input/output lines
(100 kHz repetition frequency)
¦
2 kV for power supply lines
¦
1 kV for input/output lines
(100 kHz repetition frequency)
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
¦
0.5 kV,
¦
1 kV
line(s) to line(s)
¦
0.5 kV,
¦
1 kV,
¦
2 kV
line(s) to earth
¦
0.5 kV,
¦
1 kV
line(s) to line(s)
¦
0.5 kV,
¦
1 kV,
¦
2 kV
line(s) to earth
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips, short interruptions
and voltage variations on power
supply input lines.
IEC 61000-4-11
0% U
T
for 0.5 cycle
at 0
o
, 45
o
, 90
o
,135
o
, 180
o
,
225
o
, 270
o
and 315
o
0 % U
T
for 1 cycle
at 0
o
70 % U
T
for 25/30 cycles
at 0
o
0% U
T
250/300 cycles
0% U
T
for 0.5 cycle
at 0
o
, 45
o
, 90
o
,135
o
, 180
o
,
225
o
, 270
o
and 315
o
0 % U
T
for 1 cycle
at 0
o
70 % U
T
for 25/30 cycles
at 0
o
0% U
T
250/300 cycles
Mains power quality should be
that of a typical commercial or
hospital environment. If the
user of the This X-ray System
requires continued operation
during
power
mains
interruptions,
it
is
recommended that this X-ray
System is powered from an
Uninterruptible Power Supply
or a battery.
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
or hospital environment.
NOTE - U
T
is the a.c. mains voltage prior to application of the test level.
Содержание DX-D 100
Страница 1: ...DX D 100 Mobile X Ray Unit User Manual Technical Publication 0188K EN 20201116 ...
Страница 2: ......
Страница 8: ...DX D 100 Mobile X ray Unit User Manual 0188K EN 20201116 iv ...
Страница 46: ...DX D 100 Mobile X ray Unit User Manual 38 0188K EN 20201116 This page intentionally left blank ...
Страница 80: ...DX D 100 Mobile X ray Unit User Manual 72 0188K EN 20201116 This page intentionally left blank ...
Страница 84: ...DX D 100 Mobile X ray Unit User Manual 76 0188K EN 20201116 This page intentionally left blank ...
Страница 95: ......