DD+DIS150.05E
Repair and Service
Adjustments and Calibrations
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
2.5.5
Performing a Density Reproduction Check
NOTE:
Performing this step is a verification method for step “Performing a Density
Reproduction CheckCalibration”. If the Density Reproduction Check is successful, the
internal densitometer calibration is OK.
(1) In the browser pages select menu "Service Engineer Tools - Enter Service
Command Session - 3 Print Image - 1 Test Image".
(2) Select the upper or lower tray.
(3) Print out LinCheck.zip and measure the density of the eleven 100 % to 0% fields
with a calibrated MacBeth TR-924, X-Rite 331, 341, X-Rite 301, X-Rite 310 or
Gretag D200-2 densitometer.
(4) To check linearity fill in measured values in the density reproduction check Excel
Select densitometer:
4
Dmax = 315
Square
Measured density
Target density Density diff. Check
100%
23
23
0
0
90%
50
52,2
-2,2
0
80%
78
81,4
-3,4
0
70%
105
110,6
-5,6
0
Edition 1, Revision 1
DRYSTAR 5302
Chapter 3.6 / 26
05-2009 Type
5366/100
Agfa Company Confidential
Density Reproduction =
OK
No CDM calibration needed for DT2B film
60%
135
139,8
-4,8
0
50%
162
169
-7
0
40%
193
198,2
-5,2
0
30%
221
227,4
-6,4
0
20%
250
256,6
-6,6
0
10%
280
285,8
-5,8
0
0%
308
315
-7
0
When entering the measured density values the decimal point appears in a different spot, e.g. measured value = 0.23, value to enter = 23.0
Density
Lower limit
Upper limit
0
0
0
30
-5
5
45
-7
7
60
-8
8
75
-9
9
90
-10
10
X-rite 310
105
-11
11
X-rite 331
120
-12
12
X-rite 341
135
-12
12
X-rite 301
150
-13
13
Mac Beth TR924
165
-13
13
Gretag D200-2
180
-14
14
195
-14
14
210
-14
14
225
-15
15
240
-15
15
255
-15
15
270
-16
16
285
-16
16
300
-16
16
315
-16
16
Density difference LinCheck image DT2B film in DS5302
-20
-15
-10
-5
0
5
10
15
20
0
50
100
150
200
250
300
350
Target Density
De
ns
it
y
Di
ff
e
re
n
c
e
Lower limit
Upper limit
Density difference
X-rite 341
Figure 16
(5) After the measured values are filled in, the density reproduction is calculated
automatically.
At the right hand side of the table is shown, whether the Density
Reproduction is ok or not. In case it is not ok you are requested to carry out a
CDM calibration.
Содержание Drystar 5302
Страница 2: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers ...
Страница 34: ......
Страница 64: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers ...
Страница 90: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers ...
Страница 92: ......
Страница 100: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers ...
Страница 102: ......
Страница 156: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers ...
Страница 232: ......
Страница 240: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers ...
Страница 278: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers ...
Страница 302: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers ...
Страница 364: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers ...
Страница 442: ......
Страница 453: ...DRYSTAR 5302 Type 5366 100 1st Edition r Orde no I 1 DD D S 50 05E ...
Страница 454: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers ...