DD+DIS238.06E
Generic Safety Directions
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 5
Generic Safety Directions for HealthCare Imaging Products
Page 26 of 30
Performing the electrical test according to national regulations after repair work:
WARNING:
Improper ground connections or too high leakage current may lead to electric
shocks.
After any repair work which may influence electrical safety of the product, inform the
responsible organization* about the necessity of the electrical test according to
national regulations.
If specific national regulations do not exist: It is recommended to perform the
electrical test according to IEC 62353.
Make sure, that all grounding connections are present.
*Responsible Organization:
Entity accountable for the use and maintenance of a medical equipment or a medical equipment system. The
accountable entity can be, for example, a hospital or an individual clinician.
NOTE:
Refer to the Agfa HealthCare Library for system specific IEC 62353 test documents.
Performing service activities at devices emitting laser radiation:
WARNING:
Laser radiation. Eye injury possible.
INSTRUCTION:
Strictly observe the warning notes in the service manual of devices emitting laser
radiation (See service manual chapter describing Safety Guidelines / General
Repair Instructions) and at the corresponding steps of instructions.
Strictly observe the warning labels at the modules emitting laser light. For the
meaning of the labels refer to section 3.3 in this document.
Do not look into the laser beam.
Do not open modules containing a laser. Only open modules containing a laser if
explicitly instructed to do so.
Do not keep tools in the laser beam unless explicitly instructed to do so.
Make yourself familiar with the path of the laser light and the conditions, when the
laser beam is switched on. Refer to the functional description in the corresponding
service manual.
Do not operate modules with laser outside the device.
04-2014
Agfa Company Confidential
Содержание DRYSTAR 5302 100
Страница 34: ......
Страница 64: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers...
Страница 90: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers...
Страница 92: ......
Страница 100: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers...
Страница 102: ......
Страница 156: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers...
Страница 232: ......
Страница 240: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers...
Страница 278: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers...
Страница 302: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers...
Страница 448: ......
Страница 459: ...DRYSTAR 5302 Type 5366 100 1st Edition r Orde no I 1 DD D S 50 05E...
Страница 460: ...This page is intentionally left blank to enable print out of even and odd pages on duplex printers...