118
2920B EN 20031206
Performing quality control (QC) tests
The following procedures must be performed daily, weekly or annually as
indicated.
The reason for performing quality control tests is to determine if any
significant image quality variation or deterioration has occurred which may
require corrective action. Comparing the results of the tests with the
reference values previously established does this.
This procedure allows the operator to take the necessary preventive actions
before any image quality loss can take place.
Performing the daily QC test
1
Switch on the Drystar 5300 and wait for at least 15 minutes. Refer to
‘Switching on the Drystar 5300’
2
Press the Key-operator key to enter the key-operator mode.
3
Press the Down key seven times, followed by the Confirm key to select
‘Quality control’.
A confirmation screen appears:
This test must be performed every day before any clinical film can be
processed.
1 Show settings
2 Change settings
3 Print image
4 Save configuration
5 Restore config.
6 Calibration
7 Service actions
8 Quality control
9 Installation
KO
Start QC
Procedure
YES (v)
NO (x)
QC
Содержание Drystar 5300
Страница 1: ...Drystar 5300 Reference manual POWER ...
Страница 154: ...154 2920B EN 20031206 4 Lower the film positioner unit 5 Remove the film positioner cable out of its retainers ...
Страница 156: ...156 2920B EN 20031206 8 Using the cross head screwdriver remount the 3 screws to fix the film position unit ...
Страница 160: ...160 2920B EN 20031206 ...
Страница 168: ...168 2920B EN 20031206 ...
Страница 189: ...Equipment information sheet Appendix A ...
Страница 195: ...Drystar media density response data Appendix B ...
Страница 198: ...198 2920B EN 20031206 ...
Страница 199: ...Quality Control Charts Appendix C ...
Страница 205: ...Index Appendix D ...
Страница 210: ...Printed in Belgium Published by Agfa Gevaert N V B 2640 Mortsel Belgium 2920B EN 20031206 ...